Technical Specifications
Test Principle | Impedance method、Coulter principle、Colorimetry、 Flow cytometry technology+laser light scattering、5-part differential analysis |
Parameters | 25 parameters: WBC、Lym#、Lym%、Mon#、Mon%、Eos#、Eos%、Neu#、 Neu%、Bas#、Bas%、RBC、RDW-CV、RDW-SD、HGB、MCV、MCH、MCHC、 HCT、PLT、MPV、PDW、P-LCR、P-LCC、PCT 6 research parameters: ALY#、ALY%、LIC#、LIC%、NRBC#、NRBC% 3 Histograms:WBC、RBC、PLT 3 WBC Abn Scattergrams、3D Scatter Plot |
Reagents | DIL diluent、MX-H1 lyse、MX-H2 lyse、MX-H3 lyse、CLE-P cleanser |
Sample volume | Prediluent mode: 20μL; Whole blood mode: 15μL |
Throughput | 60 samples per hour |
Display | 13.3 inch touch screen |
Data capacity | UP to 100,000 results including numeric and graphical information |
Interface | USB, LAN, HDMI; Support bi-directional LIS |
Printout | External laser/inkjet printer |
Power voltage | AC100V-240V; 50-60HZ |
Operating environment | Temperature: 10℃~30℃; Humidity: 20%~85%; Air pressure: 70kPa~106kPa |
Dimension & weight | 380mm×460mm×530mm; 42kg |
Performance
Parameter | Measuring Range | Range of relative deviation allowed (%) |
WBC | (3.5~9.5)×109/L | Not more than±15% |
RBC | (3.8~5.8)×1012/L | Not more than±6% |
HGB | (115~175) g/L | Not more than±6% |
MCV | (82~100) fL | Not more than±7% |
PLT | (125~350)×109/L | Not more than±20% |