Automatic 5-Part Hematology Analyzer MX60
Technical Specifications
Technology & Principle | |||||||||||||||||
WBC: Flow Cytometry (FCM) + Laser light scatter + Cytochemical staining method | |||||||||||||||||
RBC/PLT:Electrical impedance method | |||||||||||||||||
HGB:Non-Cyanide colorimetric method | |||||||||||||||||
Throughput | 60 samples / hour | ||||||||||||||||
Sampling modes | Whole blood, Capillary whole blood, Pre-diluted sample | ||||||||||||||||
Sample volume | Whole blood: 20μL, Capillary whole blood:20μL, Pre-diluted mode: 20μL | ||||||||||||||||
Test modes | CBC+DIFF | ||||||||||||||||
QC Mode | L-J, X-B | ||||||||||||||||
Storage Capacity | 50,000 results memory with scattergrams and histograms | ||||||||||||||||
Species of animals | Cat, dog, mouse, rabbit, horse, pig, cattle, deer, monkey, sheep, bird, chicken |
||||||||||||||||
Physical Characteristics | |||||||||||||||||
Display | 10.4 inch colour touch screen | ||||||||||||||||
Machine Dimension&Weight | Width x Height x Depth: 325mm*465mm*470mm, Weight: 30Kg | ||||||||||||||||
Package Dimension&Weight | Length x Width x Height: 630mm*490mm*630mm, Weight: 35Kg | ||||||||||||||||
Working temperature | 10-30℃ | ||||||||||||||||
Relative Humidity | 70% max | ||||||||||||||||
Reagents | |||||||||||||||||
Diluent | 20L×1 | ||||||||||||||||
Lyse | 60DIFF(500mL×1), 60LH(100mL×1) | ||||||||||||||||
Probe cleanser | 100mL×2 |
Precision | ||
Measurement | Measuring Range | Repeatability Limits Whole Blood(CV) |
WBC | 3.5-15.0x109/L | ≤ 2.0% |
RBC | 3.5-6.0x1012/L | ≤ 1.5% |
HGB | 110.0-180.0 g/L | ≤ 1.5% |
MCV | 70.0-120.0 fL | ≤ 1.0% |
PLT | 150.0-500.0x109/L | ≤ 4.0% |
Linearity | ||
Measurement | Measuring Range | Limit |
WBC | 0-100x109/L | ± 0,3 or ± 5% |
RBC | 0-8x1012/L | ± 0,05 or ± 5% |
HGB | 0-250 g/L | ± 2 or ± 2% |
PLT | 0-1000x109/L | ± 10 or ± 8% |